Comparison of Detection Results of 6 Kinds of Diagnostic Reagents for Brucellosis
CSTR:
Author:
Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    In order to compare the performance of different brucellosis diagnostic reagents in laboratory tests,In this study,96 samples of cow serum were detected by 4 methods,including rose bengal plate agglutination test (RBT),serum agglutination test (SAT),indirect enzymelinked immunosorbent assay (iELISA) and competitive enzymelinked immunosorbent assay (cELISA),and 6 different reagents.SAT results were used as the gold standard.The sensitivity,specificity and coincidence rate of the other 5 diagnostic reagents were analyzed,and the consistency of the detection results of different reagents with the same method was analyzed.The results showed that 3 positive samples and 93 negative samples were detected by SAT.RBT,iELISAA and iELISAB showed the highest sensitivity (100.00%) compared with SAT.The specificity of RBT and cELISAC was the highest (98.92%).The coincidence rates of RBT,iELISAA and cELISAC were all above 91.67%.The consistency between iELISAA and iELISAB and between cELISAC and cELISAD was poor.The results of this study provided reference for the selection of diagnostic reagents to improve the diagnostic accuracy.